PH Cube Benchtop PH – MV – Temperature Meter
20/04/2017
Perangkat pengujian pengaruh panas dan udara terhadap lapisan tipis aspal yang diputar (RTFOT)
21/04/2017Alat Uji Daktilitas Aspal
Standard Ductilometer for Ductility test – Ductimeter
Standard : EN 13398, ASTM D113, AASHTO T51, ASTM D6084
Model : 81-PV10A02
ASTM D113 : 1.1 The ductility of a bituminous material is measured by the distance to which it will elongate before breaking when two ends of a briquet specimen of the material, of the form described in Section 4, are pulled apart at a specified speed and at a specified temperature. Unless otherwise specified, the test shall be made at a temperature of 25 6 0.5°C and with a speed of 5 cm/min 6 5.0 %. At other temperatures the speed should be specified.
Referensi Standard Indonesia pengujian Daktilitas : RSNI M-04-2005 Keputusan Menteri Permukiman dan Prasarana Wilayah Nomor : 330/KPTS/M/2002
Main Features :
- 4 tension line (briquette capacity) x 1500 mm
- Easy and free access to the large testing space
- Closed-loop PID temperature control system
- Stainless steel insulated water bath
- Exclusive in-built thermoregulation system compensating the exchange of heat and cooling, resulting in a very strict temperature control, optimized by the connection to chiller (optional)
- Adjustable speed range from 5 to 100 mm/min
- High carriage return speed of 500 mm/min for greater productivity
- Elongation measurement by encoder read on display
General Descriptions :
This model fully satisfies and exceeds ASTM D113, ASTM D6084, AASHTO T51 and EN 13398 requirements. To obtain the required 25° with ± 0.02° C tolerance, circulation of cold water is necessary. A water chiller (see accessories) is ideal for this and may already be available in the laboratory but mains water can also be used. If the ambient temperature goes over 25° C, as in the tropical areas, and cold mains water is not available, the use of a water chiller is mandatory.
Water bath
Insulated stainless steel water bath with a heating system located over the entire base surface and a cooling coil (for connection to water mains or chiller) distributed over the three side walls assuring temperature uniformity without water turbulence inside the bath. All parts in contact with water are made of stainless steel.
Carriage displacement
Double-screw rods and mobile carriage enable test speeds from 5 to 100 mm/min. The system also permits the fast return of the carriage at the end of the test (500 mm/min) to reduce dead time and increase productivity. The carriage return is automatic so manual intervention is not required.
Temperature control
- A closed-loop PID system assures constant temperature within ± 0.2° C
- Plate-type base heaters give better temperature uniformity
- The thermoregulation system is based on controlling the heater (which increases the temperature of the bath) and the flow rate of the cooling coil (which decreases the temperature). This important feature assures control of the temperature within strict limits and permits the use of suitable standard water chillers
Machine control and elongation measurement
Control panel with digital display to set the testing speed: 5 to 100 mm/min, with the carriage return function. Elongation measurement by encoder.
Spesifications :
- Machine control by : Digital display panel
- Thermostatically controlled water bath temperature : at 25±0.2°C
- Temperature control system : Heater and cooling coil for connection to cold water or water chiller
- Structure : Stainless steel tank
- Briquettes capacity : 4
- Max carriage travel : 1500 mm
- Testing speed : Adjustable from 5 to 100 mm/min
- Elongation measurement by : Encoder (linear scale)
- Carriage return speed : 500 mm/min
- Power rating (approx.) : 1200 W
- Transparent cover : As optional
- Overall dimensions : (l x d x h) 2434 x 412 x 385 mm
- Weight approx. : 100 kg
Kode Pesanan :
81-B0141 Ductility mould to EN 13398
81-B0141/A Briquette mould conforming to ASTM D6084 and AASHTO T300
81-B0141/B Briquette mould conforming to ASTM D113 and AASHTO T51
81-B0141/E Briquette mould conforming to EN 13589
81-B0142 Ductility mould plate
81-PV10010 Support base for ductility machines
81-PV1002 Water chiller, flow rate 6 l/min at 1 bar, 2°C min. temperature. 1200 W cooling capacity. 230V/50-60Hz/1ph
81-PV10030 Cover for DUCTI-Meter Ductility machine
81-PV1004 Water chiller, flow rate 6 l/min at 1 bar, 2°C min. temperature. 1200 W cooling capacity. 110V/60Hz/1ph
Informasi Produk :
Yudhistira
Hp. 0811 188 6845; 0818 0816 0634
Email : naysite2@gmail.com
Reviews